6 changing calibration/validation no, 7 changing calibration/validation method, Changing calibration/validation no -14 – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 61: Changing calibration/validation method -14

<4. Display and Operations while Connected to GC8000>
4-14
IM 11B06C01-01E
4.3.6
Changing Calibration/Validation No.
To change a Calibration/Validation number, select the Calibration/Validation command from the
Measurement Status menu. The Calibration/Validation dialog box appears.
List only those for which calibration
settings have been specified.
Figure 4.17
Calibration/Validation dialog box
From the Calibration/Validation dialog box, you can select any of Calibration 1 to Calibration 6,
and any of Validation 1 to Validation 6.
If the stream number is not set, its option is not displayed. (If no stream number is set for
Calibration/Validation, its button is grayed out and cannot be selected from the dialog box.)
When you specify a Calibration/Validation number and click the Execute button, the
measurement status is changed to the Calibration/Validation of the selected calibration/validation
number.
If you select a Dummy Task (Stream 33), the Stream under Measurement is changed to the
“Replacement available.” In this case, the Dummy Task (Stream 33) is measured continuously
until you select the Start Calibration/Validation command from the Measurement Status menu.
(The chromatogram is not saved and the analysis result is not calculated.)
4.3.7
Changing Calibration/Validation Method
To change the Calibration/Validation Method, select the Calibration/Validation Method command
from the Measurement Status menu. The Calibration/Validation Method dialog box appears.
Figure 4.18
Calibration/Validation Method dialog box
5th Edition : Nov. 16, 2011-00