Appendix B List of Messages
This Appendix contains a list of error messages displayed during Engineering Terminal
operations and the countermeasures to be taken.
n
Ethernet LCD Emulator Window
Message
Meaning
Countermeasures
St
ar
t
EtherLCD cannot execute
more than 5 on a PC.
More than 5 EtherLCDs are
operated on a PC at a same time.
4 is Maximum number of
EtherLCD to be connected at a
same time.
C
on
ne
ct
io
n
Other EtherLCD connects
with the analyzer
(xxx.xxx.xxx.xxx).
The appointed IP address of the
analyzer is now connecting with
(xxx.xxx.xxx.xxx) of the EtherLCD.
An analyzer cannot communicate
with the multiple EtherLCD at a
same time. Execute the command
again after disconnecting the
current connection.
The communication has
failed, because the Analyer is
local mode.
The communication cannot be
started because the analyzer is
under local mode.
Change from local mode to
remote mode at the anlyzer, then
command again.
Cannot open communication
device.
The communication cannot be
started due to a communication
device problem.
Check the Ethernet card, setting
of TCP/IP and Operation status of
the analyzer.
Fo
rc
e
co
nn
ec
t Password is incorrect.
Force connection is not
possible.
The entered password is invalid at
the force connection.
Enter the correct password.
O
n-
Li
ne
Communication with Analyzer
xxx has been cut.
The communication has failed
between the PC and the analyzer.
Check the network between the
PC and the analyzer, then try to
connect again.
Other PC starts the force
connection.
xxx.xxx.xxx.xxx.
The communication has failed
due to the force connection by
EtherLCD of IP address xxx.xxx.
xxx.xxx.
After ending the communication
with IP address xxx.xxx.xxx.xxx,
make the connection again.
An
al
yz
er
s
el
ec
t
di
al
og
This IP address is incorrect.
Invalid IP address is entered.
Enter the correct IP address.
This IP address overlaps with
analyzer xxx.xxx.xxx. xxx.
The entered IP address is
duplicated the analyzer xxx.xxx.
xxx.xxx.
Set the different IP address.
The analyzer ID is out of
range.
Invalid Analyzer ID is entered.
Set the valid analyzer ID. The
range is 1 to 240.
5th Edition : Nov. 16, 2011-00