16 detector signal display, Detector signal display -19 – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 66

<4. Display and Operations while Connected to GC8000>
4-19
IM 11B06C01-01E
4.3.16
Detector Signal Display
To display detector signals, right-click the detector. The Detector Signal dialog box appears.
Figure 4.23
Detector Signal dialog box (The TCD at left, FID-MC at center, and FPD (similar to FID) at
right)
The display is refreshed in every second. The following items are displayed.
Table Detector signal display item
Item
Description
Type
Indicates the detector type. It is either “TCD”, “FID”, “FID-MC” (FID having methane
converter), “FPD”, or “None”.
Input Signal
The input signal is sampled every 20 msec, and this analog signal is converted into
digital form, averaged at the sampling rate, and displayed.
Filtered Signal
The input signal is filtered using the filter constant of detector signal settings and
displayed.
Standard Deviation
Indicates the standard deviation of the past 20 input signals.
Voltage
Indicates the TCD bridge voltage. (TCD only)
Current
Indicates the TCD bridge current. (TCD only)
Flame Level
Indicates the preset Flame Detection Level. (FID, FID-MC, FPD)
Therm. Signal
Displays the thermocouple signals. (FID, FID-MC, FPD)
Flame Status
Indicates the Flame Detection status (Flame On or Flame Out). (FID, FID-MC, FPD)
Voltage of MC
Indicates the converter voltage. (FID-MC only)
5th Edition : Nov. 16, 2011-00