See also, Caution, Nuploading – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 42
<3. Overview Window>
3-14
IM 11B06C01-01E
3.7
Uploading and Downloading Analyzer
Setting Parameters
When incorporating the settings specified in the Set Parameter List window of the LCD panel
into another analyzer, or when creating a backup for re-setting, you upload and download the
parameters.
Before transferring the parameters, you need to specify the ID number of the analyzer to operate.
SEE ALSO
For specifying the analyzer ID number, see section 3.6, “Selecting Analyzer.”
CAUTION
Only level-C users can upload and/or download setting parameters. Also, the operation status
must be changed to “Stop.”
n
Uploading
To upload the set parameters from the analyzer and save them to the Engineering Terminal disk,
proceed as follows:
l
Operation procedure
(1) Click on the Parameter Transfer command in the Option menu.
The submenu for transferring parameters appears.
F0321.ai
(2) Click on the Upload to PC command.
The Save As dialog box for specifying a setting parameter file appears.
F0322.ai
Enter a file name.
5th Edition : Nov. 16, 2011-00