Wave expert – Teledyne LeCroy WaveExpert 100H Operators Manual User Manual

Page 87

background image

Wave Expert

WE-OM-E Rev A

85

A Differential TDR measurement on a 1-port device is denoted by

Similarly the Differential S-parameters of a 2-port differential device are given by:

S

DD11

(Differential Port 1 Return Loss)

S

DD12

(Differential Reverse Transmission Loss)

S

DD21

(Differential Forward transmission Loss)

S

DD22

(Differential Port 2 Return Loss)

Making S-parameter Measurements

A calibration must be performed to enable S-parameter measurements selection. Refer to
“Reference Plane Calibration” for the calibration procedure.

S-parameter measurements are accessible through the Normalized trace menu:

Measurement selections under the TDR Normalization tab. Normalized trace can be displayed either

as Impedance or the corresponding S-parameter.

Depending on the type of calibration (Single-ended or Differential) and number of ports (one or two),
the corresponding s-parameters appear as selections under the TDR Normalization tab.

S-parameters are displayed in decibels (dB) over a 20 GHz frequency range, internally set by the
measurement. The trace can be adjusted using the features available in the ZOOM menu.