Operator’s manual – Teledyne LeCroy WaveExpert 100H Operators Manual User Manual
Page 188
Operator’s Manual
186
WE-OM-E Rev A
You can also enter combinations of the above units following the SI rules:
y
for the quotient of two units, the character / should be used.
y
for the product of two units, the character . should be used.
y
exponents can be represented by a digit appended to the unit without a space.
For example,
y
acceleration can be entered as M/S2 for meters per second squared.
y
volts seconds can be entered as V.S.
In some cases, the units entered may be converted to simple units. For example entering V.A will
display W (watts).
To Set Up Rescaling
1. In the menu bar, touch Math, then Math Setup... in the drop-down menu.
2. Touch a math function trace tab F1 through Fx The number of math traces available
depends on the software options loaded on your scope. See Specifications.
3. Touch
the
single function button.
4. Touch
inside
the
Source1 data entry field and select a source: channel, math trace,
memory location.
5. Touch
inside
the
Operator1 data entry field and select Rescale from the Functions
category.
6. In the dialog on the right, touch the Rescale tab.
7. Touch
inside
the
First multiply by checkbox and enter a value for a, the multiplication
factor.
8. Touch
inside
the then add: data entry field and enter a value for b, the additive constant.
9. Touch
inside
the
Override units checkbox to disregard the source waveform's units, using
the pop-up keyboard.