beautypg.com

Operator’s manual – Teledyne LeCroy WaveExpert 100H Operators Manual User Manual

Page 188

background image

Operator’s Manual

186

WE-OM-E Rev A

You can also enter combinations of the above units following the SI rules:

y

for the quotient of two units, the character / should be used.

y

for the product of two units, the character . should be used.

y

exponents can be represented by a digit appended to the unit without a space.

For example,

y

acceleration can be entered as M/S2 for meters per second squared.

y

volts seconds can be entered as V.S.

In some cases, the units entered may be converted to simple units. For example entering V.A will
display W (watts).

To Set Up Rescaling

1. In the menu bar, touch Math, then Math Setup... in the drop-down menu.
2. Touch a math function trace tab F1 through Fx The number of math traces available

depends on the software options loaded on your scope. See Specifications.

3. Touch

the

single function button.

4. Touch

inside

the

Source1 data entry field and select a source: channel, math trace,

memory location.

5. Touch

inside

the

Operator1 data entry field and select Rescale from the Functions

category.

6. In the dialog on the right, touch the Rescale tab.
7. Touch

inside

the

First multiply by checkbox and enter a value for a, the multiplication

factor.

8. Touch

inside

the then add: data entry field and enter a value for b, the additive constant.

9. Touch

inside

the

Override units checkbox to disregard the source waveform's units, using

the pop-up keyboard.