Page 78

background image

Operator’s Manual

76

WE-OM-E Rev A

Note: An easier way to adjust the value is to close the pop-up keypad and, with the “Port-“ field highlighted, turn the front
panel

A

DJUST knob to set the value.

3. In the manual deskew mode, adjust the “Channel Skew” to obtain a reading of 0 ps. Touch

Next> when the Skew is 0 ps.

4. The next step in the manual deskew procedure requires you to remove the cable

connecting the two modules, and to connect to the modules the two cables that will be
connected to the DUT. The free ends of the cables will be left unconnected at this time.
Touch Next> when ready.