Operator’s Manual
170
WE-OM-E Rev A
the power at a specific
frequency of a waveform
(narrow band).
and XMAP options.
Standard in
DDA-5005A.
N-cycle jitter
Peak-to-peak jitter between
edges spaced n UI apart.
Compares the expected
time to the actual time
of leading edges n bits
apart.
Available in SDA
analyzers.
NLTS
Provides a measurement of
the nonlinear transition shift
for a prml signal.
Available with DDM2
option.
Standard in
DDA-5005AA.
Npts
Number of points in the
waveform between the
cursors.
Standard parameter.
Overshoot-
Overshoot negative: Amount
of overshoot following a
falling edge, as percentage
of amplitude.
(base - min.)/ampl x 100 Waveform must contain
at least one falling
edge. On signals not
having two major levels
(triangle or saw-tooth
waves, for example),
may not give
predictable results.
Standard parameter.
Overshoot+
Overshoot positive: Amount
of overshoot following a
rising edge specified as
percentage of amplitude.
(max. - top)/ampl x 100 Waveform must contain
at least one rising edge.
On signals not having
two major levels
(triangle or saw-tooth
waves, for example),
may not give
predictable results.
Standard parameter.
Overwrite
Ratio of residual-to-original
power of a low frequency
waveform overwritten by a
higher frequency.
Available with DDM2
option.
Standard in
DDA-5005A.
Param Script
Visual Basic or Java script
that produces a
measurement from one or
Available with XMAP
option.