Operator’s Manual
214
WE-OM-E Rev A
5. Touch
inside
the
Pass If field, and select a Boolean condition from the pop-up menu.
6. If you want to set up the test to end after a finite number of sweeps, touch the Stop Test
checkbox. Then touch inside the After data entry field and enter a value, using the pop-up
numeric keypad.
7. Under "If", touch either the Pass
or Fail
button to set the actions to occur
upon your waveform's passing or failing the test.
8. Under "Then", touch the actions you want to occur: stop test, sound alarm, print result, emit
pulse, or save the waveform. If you want to have the results printed and your scope is not
equipped with a printer, be sure that the it is connected to a local or network printer. See
Printing.
9. If you want to save your waveform automatically, touch the Save Setup. This will take you
out of the current dialog and will open the "Save Waveform" dialog. See Saving and
Recalling Waveforms.
10. Test your Pass/Fail conditions by touching the Force Actions Once button. Press the
Clear All button to quickly uncheck all checkboxes if you want to change your selections.
Comparing a Single Parameter
1. Touch
Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.
2. Touch
a
Qx tab; a setup dialog for that position will open.
3. Touch
inside
the
Source1 field and select a source from the pop-up menu.
4. Touch
inside
the
Condition field in the main dialog and select ParamCompare
.
5. Touch
inside
the
Compare Values field and select All or Any from the pop-up menu
. By selecting All, the test will be true only if every waveform falls within the limit
that you will set. By selecting Any, the test will be true if just one waveform falls within the
limit.
6. Touch
inside
the
Condition field in the "ParamCompare" mini-dialog and select a math
operator from the pop-up menu: