Operator’s Manual

214

WE-OM-E Rev A

5. Touch

inside

the

Pass If field, and select a Boolean condition from the pop-up menu.

6. If you want to set up the test to end after a finite number of sweeps, touch the Stop Test

checkbox. Then touch inside the After data entry field and enter a value, using the pop-up
numeric keypad.

7. Under "If", touch either the Pass

or Fail

button to set the actions to occur

upon your waveform's passing or failing the test.

8. Under "Then", touch the actions you want to occur: stop test, sound alarm, print result, emit

pulse, or save the waveform. If you want to have the results printed and your scope is not
equipped with a printer, be sure that the it is connected to a local or network printer. See
Printing.

9. If you want to save your waveform automatically, touch the Save Setup. This will take you

out of the current dialog and will open the "Save Waveform" dialog. See Saving and
Recalling Waveforms.

10. Test your Pass/Fail conditions by touching the Force Actions Once button. Press the

Clear All button to quickly uncheck all checkboxes if you want to change your selections.

Comparing a Single Parameter

1. Touch

Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.

2. Touch

a

Qx tab; a setup dialog for that position will open.

3. Touch

inside

the

Source1 field and select a source from the pop-up menu.

4. Touch

inside

the

Condition field in the main dialog and select ParamCompare

.

5. Touch

inside

the

Compare Values field and select All or Any from the pop-up menu

. By selecting All, the test will be true only if every waveform falls within the limit

that you will set. By selecting Any, the test will be true if just one waveform falls within the
limit.

6. Touch

inside

the

Condition field in the "ParamCompare" mini-dialog and select a math

operator from the pop-up menu: