Delay Parameter
(Find Range selected)

Waveform Gated Parameters

To Set Up Waveform Qualifiers

1. From the menu bar, select Measure, then Measure Setup... from the drop-down menu.
2. Touch

a

Px tab to open the setup dialog.

3. Touch

inside

the

Source field and select a source from the pop-up menu.

4. Touch

inside

the

Measure field and select a parameter from the pop-up menu.

5. Touch

the

Accept tab of the right-hand dialog, then touch the Values Based on

Waveform State checkbox.

6. Touch

inside

the

When Wform field and select the gating source.

7. Touch

inside

the

State Is field and select High or Low from the pop-up menu. Parameter

measurements on the subject waveform will only be taken when the gating waveform is in
the selected state.

8. Touch inside the Level Type field and select Absolute or Percent from the pop-up menu.
9. Touch

inside

the

Level field and enter the crossing level value at which you want

measurements to begin. Alternatively, touch the Find Level button to automatically select
the 50% level of your gating waveform.