Teledyne LeCroy WaveExpert 100H Operators Manual User Manual
Page 28

Operator’s Manual
26
WE-OM-E Rev A
SPECIFICATIONS
Note: Specifications are subject to change without notice.
Sampling Heads
Module Performance (Guaranteed Values)
Module
Rise time RMS noise
Aberrations
Bandwidth
ST-20
18 ps
700 µV
First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%
20 GHz
SE-30
12 ps
1 mV
First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%
30 GHz
SE-50
8 ps
2 mV
First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%
50 GHz
SE-70
5 ps
3 mV
First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%
70 GHz*
SE-100
4 ps
3 mV
First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%
100 GHz*
* limited by connector molding
ME-15 Module Extender: 22 ps
Vertical System
Maximum Input Channels: 4
Maximum Input Range: +/- 2 V
Dynamic Range: 2 V
p-p
Input Impedance: 50 ohms ±1%
Input Coupling: DC
Max Input Voltage: +/- 2.5 V
Static Sensitivity: The sampling module inputs are highly static sensitive. A grounding strap
must be worn at all times when handling the modules.
Installation (Overvoltage) Category: CAT I