beautypg.com

Teledyne LeCroy WaveExpert 100H Operators Manual User Manual

Page 28

background image

Operator’s Manual

26

WE-OM-E Rev A

SPECIFICATIONS

Note: Specifications are subject to change without notice.

Sampling Heads

Module Performance (Guaranteed Values)

Module

Rise time RMS noise

Aberrations

Bandwidth

ST-20

18 ps

700 µV

First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%

20 GHz

SE-30

12 ps

1 mV

First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%

30 GHz

SE-50

8 ps

2 mV

First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%

50 GHz

SE-70

5 ps

3 mV

First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%

70 GHz*

SE-100

4 ps

3 mV

First 40 ps +/-10%
40 ps to 200 ps +/-5%
200 ps to 10 ns +/-2%

100 GHz*

* limited by connector molding

ME-15 Module Extender: 22 ps

Vertical System

Maximum Input Channels: 4

Maximum Input Range: +/- 2 V

Dynamic Range: 2 V

p-p

Input Impedance: 50 ohms ±1%

Input Coupling: DC

Max Input Voltage: +/- 2.5 V

Static Sensitivity: The sampling module inputs are highly static sensitive. A grounding strap

must be worn at all times when handling the modules.

Installation (Overvoltage) Category: CAT I