Wave Expert

WE-OM-E Rev A

67

1. In the menu bar, touch Vertical, then select a channel from the drop-down menu.

2. Touch

the

Dark Cal button

.

3. A dialog box will appear instructing you to disconnect any optical signal from the channel.

Touch the Start Dark Calibration button when this is done.

4. Dark calibration will be performed, and the dark level will appear in the Dark Cal Level

field.