Teledyne LeCroy WaveExpert 100H Operators Manual User Manual
Page 33
![background image](/manuals/353451/33/background.png)
Wave Expert
WE-OM-E Rev A
31
100 to 120 V
rms
(±10%) at 400 Hz (±5%); Automatic AC voltage selection
Voltage Range:
90 to 264 V
rms
90 to 132 V
rms
Frequency Range:
47 to 63 Hz
380 to 420 Hz
Power Consumption: On State: 400 watts (400 VA) depending on accessories installed
(sampling modules, PC port plug-ins, etc.). Standby State: 12 watts
Fuse: One 5x20 mm fuse (T6.3 A/250 V)
Battery Backup: Front panel settings retained for two years minimum
Physical Dimensions (HWD): 264 mm x 397 mm x 491 mm (10.4 in. x 15.6 in. x 19.3 in.); height
measurement excludes foot pads
Weight: 16 kg (36 lbs.)
Shipping Weight: 22 kg (48 lbs.)
Warranty and Service
1-year warranty; calibration recommended yearly
Optional service programs include extended warranty, upgrades, and calibration services.
Environmental Characteristics
Temperature
Operating: 5 to 40 °C
Storage (non-operating): -20 to +60 °C
Humidity
Operating: Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to
50% relative humidity at 40 °C
Storage (non-operating): 5 to 95% RH (non-condensing) as tested per MIL-PRF-28800F
Altitude
Operating: Up to 2000 m
Storage (non-operating): Up to 12,192 m (40,000 ft)
Random Vibration
Operating: 0.31 grms, 5 Hz to 500 Hz, 15 minutes in each of 3 orthogonal axes
Non-operating: 2.4 grms, 5 to 500 Hz, 15 minutes in each of 3 orthogonal axes
Shock
Functional Shock: 20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of 3
orthogonal axes, 18 shocks total