Teledyne LeCroy WaveExpert 100H Operators Manual User Manual
Page 218
![background image](/manuals/353451/218/background.png)
Operator’s Manual
216
WE-OM-E Rev A
.
8. Touch
inside
the
Limit field and enter a value, using the pop-up numeric keypad. This
value takes the dimension of the parameter that you are testing. For example, if you are
testing a time parameter, the unit is seconds.
9. If you chose either WithinDeltaPct
or WithinDeltaAbs
from the
Condition menu, touch inside the Delta field and enter a value.
Mask Testing
1. Touch
Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.
2. Touch
a
Qx tab; a setup dialog for that position will open.
3. Touch
inside
the
Source1 field and select a source from the pop-up menu.
4. Touch
inside
the
Condition field in the main dialog and select Mask Test
.
5. From the "Test" mini-dialog, make a selection in the Test is True when group of buttons
.
This selection means, for example, that if you select All In the test will be False if even a
single waveform falls outside the mask.
6. From
Show Markers, choose whether or not to have mask violations displayed.
7. If you are loading a pre-existing mask, touch the Load Mask tab, then the File button. You
can then enter the file name or browse to its location.
8. If you want to make a mask from your waveform, touch the Make Mask tab.