Teledyne LeCroy WaveExpert 100H Operators Manual User Manual

Page 83

background image

Wave Expert

WE-OM-E Rev A

81

Differential), number of ports (one or two), and the calibration method (SL(T) or OSL(T)).

5. Adjust the horizontal position knob such that the second step reflection is within the first

two horizontal divisions. Adjust the horizontal scale and position to display the entire
response of the DUT. To ensure the response of the DUT is captured, connect a Short to
the end of the DUT and adjust the timebase such that the Short measurement is visible
within the window of interest. Press the Set Default Timebase Settings control to preset
the resolution and time length to the default settings. These settings are optimum for a
broad range of devices. The default settings are : 40 Ks, 2 ns/div, and 500 fs/sample.

6. Select the Calibration kit that will be used for the calibration. A custom calibration kit can be

created for PCB-based standards by entering in the coefficients for the Short, Open, and
Load standards.