beautypg.com

Preset:user[]:set:otp, Preset:user[]:set:otp -73 – RIGOL DP832A User Manual

Page 97

background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-73

:PRESet:USER[]:SET:OTP

Syntax

:PRESet:USER[]:SET:OTP {ON|OFF}
:PRESet:USER[]:SET:OTP?

Description

Enable or disable the over-temperature protection (OTP) function of the specified

user-defined setting.
Query the status of the over-temperature protection (OTP) function of the specified

user-defined setting.

Parameter

Name

Type

Range

Default

[]

Discrete

1|2|3|4

None

{ON|OFF}

Bool

ON|OFF

ON

Explanation

When [] is 1, 2, 3 or 4, the system sets the related parameters of the first,

second, third or fourth group of user-defined setting respectively. When [] is

omitted, the system sets the parameters of the user-defined setting currently

selected by default.

Return Format

The query returns ON or OFF.

Example

:PRES:USER1:SET:OTP ON /*Enable the OTP function of the first group of the

user-defined setting*/

:PRES:USER1:SET:OTP? /*Query the status of the OTP function of the first

group of the user-defined setting and the query

returns ON*/

This manual is related to the following products: