Source[]]:voltage:protection[:level, Source[]]:voltage:protection[:level] -94 – RIGOL DP832A User Manual

Page 118

background image

RIGOL

Chapter 2 Command System

2-94

DP800 Programming Guide

[:SOURce[]]:VOLTage:PROTection[:LEVel]

Syntax

[:SOURce[]]:VOLTage:PROTection[:LEVel] {|MINimum|MAXimum}
[:SOURce[]]:VOLTage:PROTection[:LEVel]? [{MINimum|MAXimum}]

Description

Set the overvoltage protection (OVP) value of the specified channel.
Query the overvoltage protection (OVP) value of the specified channel.

Parameter

Name

Type

Range

Default

[]

Integer

1|2|3

1

Real

Refer to Table 2-2

Explanation

 When the overvoltage protection function of the specified channel is enabled

(

[:SOURce[]]:VOLTage:PROTection:STATe

), the output turns off

automatically when the output voltage exceeds the overvoltage protection limit

currently set. You can send the

[:SOURce[]]:VOLTage:PROTection:TRIPped?

command to query whether

overvoltage protection occurred on the specified channel currently.

 When [:SOURce[]] or [] is omitted, the system sets the

corresponding parameter of the channel currently selected.

 You can also send the

:OUTPut:OVP:VALue

command to set the OVP value of

the specified channel.

Return Format

The query returns the overvoltage protection value of the specified channel, for

example, 8.800.

Example

:VOLT:PROT 8.8 /*Set the OVP value of the current channel to 8.8V*/
:VOLT:PROT? /*Query the OVP value of the current channel and the query

returns 8.800*/

Related

Commands

[:SOURce[]]:VOLTage:PROTection:STATe
[:SOURce[]]:VOLTage:PROTection:TRIPped?
:OUTPut:OVP:VALue