Output:range, Output:range -62 – RIGOL DP832A User Manual
Page 86
RIGOL
2-62
DP800 Programming Guide
:OUTPut:RANGe
Syntax
:OUTPut:RANGe {P20V|P40V|LOW|HIGH}
:OUTPut:RANGe?
Description
Select the current range of the channel.
Query the range currently selected of the channel.
Parameter
Name
Type
Range
Default
{P20V|P40V|LOW|HIGH}
Discrete
P20V|P40V|LOW|HIGH
P20V
Explanation
This command is only applicable to single-channel model (DP811A/DP811).
P20V or LOW represents Range1: 20V/10A and P40V or HIGH represents
Range2: 40V/5A.
the current range.
Return Format
The query returns the rated voltage and current values of the range selected, for
example, 20V/10A。
Example
:OUTP:RANG P20V /*Select Range1: 20V/10A as the current range*/
:OUTP:RANG? /*Query the range currently selected and the query returns
20V/10A*/
:OUTP:RANG HIGH /*Select Range2: 40V/5A as the current range*/
:OUTP:RANG? /*Query the range currently selected and the query returns
40V/5A*/
Related
Command