System:klock:state, System:language:type, System:klock:state -120 – RIGOL DP832A User Manual
Page 144: System:language:type -120
RIGOL
2-120
DP800 Programming Guide
:SYSTem:KLOCk:STATe
Syntax
:SYSTem:KLOCk:STATe {ON|OFF}
:SYSTem:KLOCk:STATe?
Description
Enable or disable the remote lock.
Query the status of the remote lock.
Parameter
Name
Type
Range
Default
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
When the instrument is in remote lock mode, all the keys at the front panel
(except the output switch key
of each channel,
and the power
switch key
) are not available and
is displayed in the status bar on the
user interface (
is not displayed).
command to enable or disable the
remote lock.
Return Format
The query returns ON or OFF.
Example
:SYST:KLOC:STAT ON /*Enable the remote lock*/
:SYST:KLOC:STAT? /*Query the status of the remote lock and the query
returns ON*/
Related
Command
:SYSTem:LANGuage:TYPE
Syntax
:SYSTem:LANGuage:TYPE {EN|CH|JAP|KOR|GER|POR|POL|CHT|RUS}
:SYSTem:LANGuage:TYPE?
Description
Set the system language.
Query the current system language type.
Parameter
Name
Type
Range
Default
{EN|CH|JAP|
KOR|GER|POR|
POL|CHT|RUS}
Discrete
EN|CH|JAP|
KOR|GER|POR|
POL|CHT|RUS
CH
Return Format
The query returns English, Chinese, Japanese, Korean, German, Portuguese, Polish,
Chinese(T) or Russian.
Example
:SYST:LANG:TYPE EN /*Set the system language to English*/
:SYST:LANG:TYPE? /*Query the current system language type and the query
returns English*/