beautypg.com

System:klock:state, System:language:type, System:klock:state -120 – RIGOL DP832A User Manual

Page 144: System:language:type -120

background image

RIGOL

Chapter 2 Command System

2-120

DP800 Programming Guide

:SYSTem:KLOCk:STATe

Syntax

:SYSTem:KLOCk:STATe {ON|OFF}

:SYSTem:KLOCk:STATe?

Description

Enable or disable the remote lock.
Query the status of the remote lock.

Parameter

Name

Type

Range

Default

{ON|OFF}

Bool

ON|OFF

OFF

Explanation

 When the instrument is in remote lock mode, all the keys at the front panel

(except the output switch key

of each channel,

and the power

switch key

) are not available and

is displayed in the status bar on the

user interface (

is not displayed).

You can also send the

:SYSTem:RWLock

command to enable or disable the

remote lock.

Return Format

The query returns ON or OFF.

Example

:SYST:KLOC:STAT ON /*Enable the remote lock*/
:SYST:KLOC:STAT? /*Query the status of the remote lock and the query

returns ON*/

Related

Command

:SYSTem:RWLock

:SYSTem:LANGuage:TYPE

Syntax

:SYSTem:LANGuage:TYPE {EN|CH|JAP|KOR|GER|POR|POL|CHT|RUS}
:SYSTem:LANGuage:TYPE?

Description

Set the system language.
Query the current system language type.

Parameter

Name

Type

Range

Default

{EN|CH|JAP|

KOR|GER|POR|

POL|CHT|RUS}

Discrete

EN|CH|JAP|

KOR|GER|POR|

POL|CHT|RUS

CH

Return Format

The query returns English, Chinese, Japanese, Korean, German, Portuguese, Polish,

Chinese(T) or Russian.

Example

:SYST:LANG:TYPE EN /*Set the system language to English*/
:SYST:LANG:TYPE? /*Query the current system language type and the query

returns English*/

This manual is related to the following products: