beautypg.com

Analyzer:endtime, Analyzer:file, Analyzer:endtime -4 – RIGOL DP832A User Manual

Page 28: Analyzer:file? -4

background image

RIGOL

Chapter 2 Command System

2-4

DP800 Programming Guide

:ANALyzer:ENDTime

Syntax

:ANALyzer:ENDTime {|MINimum|MAXimum}
:ANALyzer:ENDTime? [MINimum|MAXimum]

Description

Set the end time of the analyzer.
Query the end time of the analyzer.

Parameter

Name

Type

Range

Default

Integer

Refer to the "Explanation"

Explanation

 You can only set the end time when valid record file is opened (refer to

the

:ANALyzer:FILE?

command).

 When the groups of the record file opened is less than or equal to 2048, the

range of the end time is from the start time to the maximum record time

(groups times record period) of the record file opened and the default is the

maximum record time of the record file opened.

 When the groups of the record file opened is greater than 2048, the range of

the end time is from the start time to the product of the record period of the

record file opened times 2048 and the default is the product of the record

period of the record file opened times 2048.

 When receiving the

:ANALyzer:ANALyze

command, the analyzer will analyze

the recorded data between the start time and end time.

Return Format

The query returns an integer, for example, 125.

Example

:ANAL:ENDT 125 /*Set the end time of the analyzer to 125s*/
:ANAL:ENDT? /*Query the current end time and the query returns 125*/

Related

Commands

:ANALyzer:ANALyze
:ANALyzer:FILE?
:ANALyzer:STARTTime

:ANALyzer:FILE?

Syntax

:ANALyzer:FILE?

Description

Query the record file currently opened.

Return Format

When valid record file is currently opened, the query returns the directory of the file

currently opened, for example, C:\REC 10:test.ROF; when no valid record file is

currently opened, the query returns NULL.

This manual is related to the following products: