Chapter 2 Command System

2-8

DP800 Programming Guide

:APPLy Command

The :APPLy command provides the most straightforward method to program the power supply over the

remote interface. For the multi-channel models, you can select the specified channel and set the voltage

and current in a single command; for the single-channel model, you can set the voltage and current in a

single command. When the setting values are within the parameter ranges of the corresponding channel

(multi-channel models) or range (single-channel model) of the specified model, the output voltage and

current will change to the setting values immediately after executing this command. The voltage/current

ranges and default values of each channel (multi-channel models) or range (single-channel model) of

different models are as shown in the table below.

Table 2-1 Voltage/current ranges and default values of each channel (range) of different models of DP800

series

Channel (Range)

Voltage/Current

Settable Range

Voltage/Current

Default Value

DP831A

CH1 (8V/5A)

CH2 (30V/2A)

CH3 (-30V/2A)

0V to 8.4V/0A to 5.3A

0V to 32V/0A to 2.1A

0V to -32V/0A to 2.1A

0.000V/5.0000A

00.000V/2.0000A

00.000V/2.0000A

DP832A

CH1 (30V/3A)

CH2 (30V/3A)

CH3 (5V/3A)

0V to 32V/0A to 3.2A

0V to 32V/0A to 3.2A

0V to -5.3V/0A to 3.2A

00.000V/3.000A

00.000V/3.000A

0.000V/3.000A

DP821A

CH1 (60V/1A)

CH2 (8V/10A)

0V to 63V/0A to 1.05A

0V to 8.4V/0A to 10.5A

00.000V/1.0000A

0.000V/10.000A

DP811A

Range1 (20V/10A)

Range2 (40V/5A)

0V to 21V/0A to 10.5A

0V to 42V/0A to 5.3A

00.000V/05.0000A

DP831

[1]

CH1 (8V/5A)

CH2 (30V/2A)

CH3 (-30V/2A)

0V to 8.4V/0A to 5.3A

0V to 32V/0A to 2.1A

0V to -32V/0A to 2.1A

0.000V/5.000A

00.00V/2.000A

00.00V/2.000A

DP832

[1]

CH1 (30V/3A)

CH2 (30V/3A)

CH3 (5V/3A)

0V to 32V/0A to 3.2A

0V to 32V/0A to 3.2A

0V to -5.3V/0A to 3.2A

00.00V/3.000A

00.00V/3.000A

0.00V/3.000A

DP821

[1]

CH1 (60V/1A)

CH2 (8V/10A)

0V to 63V/0A to 1.05A

0V to 8.4V/0A to 10.5A

00.00V/1.000A

0.00V/10.00A

DP811

[1]

Range1 (20V/10A)

Range2 (40V/5A)

0V to 21V/0A to 10.5A

0V to 42V/0A to 5.3A

00.00V/05.00A

Note

[1]

: When DP831 (DP832, DP821 or DP811) is installed with the high resolution option, its voltage/current settable

range and default value of each channel are the same with those of DP831A (DP832A, DP821A or DP811A).

This manual is related to the following products: