Apply command, Apply command -8 – RIGOL DP832A User Manual
Page 32
RIGOL
2-8
DP800 Programming Guide
:APPLy Command
The :APPLy command provides the most straightforward method to program the power supply over the
remote interface. For the multi-channel models, you can select the specified channel and set the voltage
and current in a single command; for the single-channel model, you can set the voltage and current in a
single command. When the setting values are within the parameter ranges of the corresponding channel
(multi-channel models) or range (single-channel model) of the specified model, the output voltage and
current will change to the setting values immediately after executing this command. The voltage/current
ranges and default values of each channel (multi-channel models) or range (single-channel model) of
different models are as shown in the table below.
Table 2-1 Voltage/current ranges and default values of each channel (range) of different models of DP800
series
Channel (Range)
Voltage/Current
Settable Range
Voltage/Current
Default Value
DP831A
CH1 (8V/5A)
CH2 (30V/2A)
CH3 (-30V/2A)
0V to 8.4V/0A to 5.3A
0V to 32V/0A to 2.1A
0V to -32V/0A to 2.1A
0.000V/5.0000A
00.000V/2.0000A
00.000V/2.0000A
DP832A
CH1 (30V/3A)
CH2 (30V/3A)
CH3 (5V/3A)
0V to 32V/0A to 3.2A
0V to 32V/0A to 3.2A
0V to -5.3V/0A to 3.2A
00.000V/3.000A
00.000V/3.000A
0.000V/3.000A
DP821A
CH1 (60V/1A)
CH2 (8V/10A)
0V to 63V/0A to 1.05A
0V to 8.4V/0A to 10.5A
00.000V/1.0000A
0.000V/10.000A
DP811A
Range1 (20V/10A)
Range2 (40V/5A)
0V to 21V/0A to 10.5A
0V to 42V/0A to 5.3A
00.000V/05.0000A
DP831
[1]
CH1 (8V/5A)
CH2 (30V/2A)
CH3 (-30V/2A)
0V to 8.4V/0A to 5.3A
0V to 32V/0A to 2.1A
0V to -32V/0A to 2.1A
0.000V/5.000A
00.00V/2.000A
00.00V/2.000A
DP832
[1]
CH1 (30V/3A)
CH2 (30V/3A)
CH3 (5V/3A)
0V to 32V/0A to 3.2A
0V to 32V/0A to 3.2A
0V to -5.3V/0A to 3.2A
00.00V/3.000A
00.00V/3.000A
0.00V/3.000A
DP821
[1]
CH1 (60V/1A)
CH2 (8V/10A)
0V to 63V/0A to 1.05A
0V to 8.4V/0A to 10.5A
00.00V/1.000A
0.00V/10.00A
DP811
[1]
Range1 (20V/10A)
Range2 (40V/5A)
0V to 21V/0A to 10.5A
0V to 42V/0A to 5.3A
00.00V/05.00A
Note
[1]
: When DP831 (DP832, DP821 or DP811) is installed with the high resolution option, its voltage/current settable
range and default value of each channel are the same with those of DP831A (DP832A, DP821A or DP811A).