2-99
:STATus:QUEStionable:ENABle
Syntax
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:ENABle?
Description
Enable the bits in the enable register of the questionable status register.
Query the bits currently enabled in the enable register of the questionable status
register.
Parameter
Name
Type
Range
Default
Discrete
Refer to the "Explanation"
None
Explanation
weights of the bits to be enabled in the enable register of the questionable
status register. For the definitions of the bits in the questionable status
registers of the multi-channel models and single-channel model and their
corresponding decimal values, refer to Table 1-1 and Table 1-4.
Take the single-channel model (DP811A) as an example. To enable bit0
(Voltage, CC mode) and bit4 (Overtemperature) in the enable register of the
questionable status register, set
2
0
+2
4
=17).
After the bits in the enable register of the questionable status register are
enabled, the system reports the states of the corresponding bits to the status
byte register.
When
register of the questionable status register.
Return Format
The query returns a decimal value corresponding to the sum of the binary weights
of the bits to be enabled in the enable register of the questionable status register,
for example, 17.
Example
:STAT:QUES:ENAB 17 /*Enable bit0 and bit4 in the enable register of the
questionable status register*/