Output:ocp:clear, Output:ocp:clear -55 – RIGOL DP832A User Manual
Page 79
RIGOL
DP800 Programming Guide
2-55
:OUTPut:OCP:CLEAR
Syntax
:OUTPut:OCP:CLEAR [CH1|CH2|CH3]
Description
Clear the label of the overcurrent protection occurred on the specified channel.
Parameter
Name
Type
Range
Default
[CH1|CH2|CH3]
Discrete
CH1|CH2|CH3
None
Explanation
Before executing the command, make sure that the reason that causes the
OCP on the specified channel is cleared (you can decrease the output current
to be below the OCP value or increase the OCP value to be greater than the
output current).
When [CH1|CH2|CH3] is omitted, the system clears the label of the OCP
occurred on the current channel.
[:SOURce[
to clear the circuit and label of the OCP occured on the specified channel and
turn on the corresponding channel output.
query whether OCP occurred on the specified channel.
Example
:OUTP:OCP:QUES? CH1 /*Query whether overcurrent protection currently
occurred on CH1 and the query returns YES*/
(Operation: decrease the output current to be below the OCP value or increase the
OCP value to be greater than the output current)
:OUTP:OCP:CLEAR CH1 /*Clear the label of the overcurrent protection occurred
on CH1*/
:OUTP:OCP:QUES? CH1 /*Query whether overcurrent protection occurred on
CH1 and the query returns NO*/
Related
Commands
:OUTPut:OCP:ALAR?
:OUTPut:OCP:QUES?
[:SOURce[