System:onoffsync, System:otp, System:onoffsync -122 – RIGOL DP832A User Manual
Page 146: System:otp -122
RIGOL
2-122
DP800 Programming Guide
:SYSTem:ONOFFSync
Syntax
:SYSTem:ONOFFSync {ON|OFF}
:SYSTem:ONOFFSync?
Description
Turn on or off the on/off sync function.
Query whether the on/off sync function is turned on.
Parameter
Name
Type
Range
Default
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
The specified channels of DP831A, DP831, DP832A and DP832 support the track
function; you can turn on or off the on/off sync function according to your need.
When the on/off sync function is turned on, the system will track the output
status of the channel; when the on/off sync function is turned off, the system will
not track the output status of the channel.
Return Format
The query returns ON or OFF.
Example
:SYST:ONOFFS ON /*Turn on the on/off sync function*/
:SYST:ONOFFS? /*Query whether the on/off sync function is turned on and
the query returns ON*/
Related
Command
:SYSTem:OTP
Syntax
:SYSTem:OTP {ON|OFF}
:SYSTem:OTP?
Description
Enable or disable the over-temperature protection (OTP) function.
Query the status of the over-temperature protection function.
Parameter
Name
Type
Range
Default
{ON|OFF}
Bool
ON|OFF
ON
Explanation
When the OTP function is enabled, the instrument turns off the output
automatically when the temperature inside the instrument reaches the limit.
Return Format
The query returns ON or OFF.
Example
:SYST:OTP ON /*Enable the OTP function*/
:SYST:OTP? /*Query the status of the OTP function and the query returns ON*/