beautypg.com

Timer:templet:riserate, Timer:templet:select, Timer:templet:riserate -140 – RIGOL DP832A User Manual

Page 164: Timer:templet:select -140

background image

RIGOL

Chapter 2 Command System

2-140

DP800 Programming Guide

:TIMEr:TEMPlet:RISERate

Syntax

:TIMEr:TEMPlet:RISERate
:TIMEr:TEMPlet:RISERate?

Description

Set the rise index of ExpRise.
Query the rise index of ExpRise.

Parameter

Name

Type

Range

Default

Integer

0 to 10

0

Explanation

When the templet currently selected is ExpRise (

:TIMEr:TEMPlet:SELect

), the timer

parameters created cannot reach the maximum due to the characteristic of the

exponential function. The range of the timer parameters created is related to the

rise index currently set. The larger the rise index is, the larger the range of the timer

parameters will be.

Return Format

The query returns an integer from 0 to 10, for example, 5.

Example

:TIME:TEMP:RISER 5 /*Set the rise index of ExpRise to 5*/
:TIME:TEMP:RISER? /*Query the rise index of ExpRise and the query returns

5*/

Related

Command

:TIMEr:TEMPlet:SELect

:TIMEr:TEMPlet:SELect

Syntax

:TIMEr:TEMPlet:SELect {SINE|SQUARE|RAMP|UP|DN|UPDN|RISE|FALL}
:TIMEr:TEMPlet:SELect?

Description

Select the desired templet type.
Query the templet type currently selected.

Parameter

Name

Type

Range

Default

{SINE|SQUARE|RAMP|

UP|DN|UPDN|RISE|FALL}

Discrete

SINE|SQUARE|RAMP|

UP|DN|UPDN|RISE|FALL

SINE

Return Format

The query returns SINE, SQUARE, RAMP, UP, DN, UPDN, RISE or FALL.

Example

:TIME:TEMP:SEL UP /*Select the UP templet*/
:TIME:TEMP:SEL? /*Query the templet type currently selected and the query

returns UP*/

This manual is related to the following products: