System:saver, System:self:test:board, System:saver -125 – RIGOL DP832A User Manual
Page 149: System:self:test:board? -125
RIGOL
DP800 Programming Guide
2-125
:SYSTem:SAVer
Syntax
:SYSTem:SAVer {ON|OFF}
:SYSTem:SAVer?
Description
Enable or disable the screen saver function.
Query the status of the screen saver function.
Parameter
Name
Type
Range
Default
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
When the screen saver function is enabled, the instrument will enter the screen
saver mode automatically after standing by for 25 minutes and will enter the black
screen state after another 12.5 minutes.
Return Format
The query returns ON or OFF.
Example
:SYST:SAV ON /*Enable the screen saver function*/
:SYST:SAV? /*Query the status of the screen saver function and the query
returns ON*/
:SYSTem:SELF:TEST:BOARD?
Syntax
:SYSTem:SELF:TEST:BOARD? [{TOP|BOTTOM}]
Description
Query the self-test results of TopBoard and BottomBoard.
Parameter
Name
Type
Range
Default
{TOP|BOTTOM}
Discrete
TOP|BOTTOM
None
Explanation
TOP: query the self-test result of TopBoard; BOTTOM: query the self-test result
of BottomBoard; when the parameter {TOP|BOTTOM} is omitted, the
command queries the self-test results of TopBoard and BottomBoard at the
same time.
command to query the self-test results of the
instrument (including the self-test results of the TopBoard, BottomBoard and
fan).
Return Format
The query returns PASS or FAIL. When the parameter {TOP|BOTTOM} is omitted,
the query returns the self-test results of TopBoard and BottomBoard (separated by
comma) at the same time, for example, PASS,PASS.
Example
:SYST:SELF:TEST:BOARD? /*Query the self-test results of TopBoard and
BottomBoard and the query returns PASS,PASS*/
Related
Command