Output:ovp:clear, Output:ovp:clear -59 – RIGOL DP832A User Manual
Page 83
RIGOL
DP800 Programming Guide
2-59
:OUTPut:OVP:CLEAR
Syntax
:OUTPut:OVP:CLEAR [CH1|CH2|CH3]
Description
Clear the label of the overvoltage protection occurred on the specified channel.
Parameter
Name
Type
Range
Default
[CH1|CH2|CH3]
Discrete
CH1|CH2|CH3
None
Explanation
Before executing the command, make sure that the reason that causes the
OVP on the specified channel is cleared (you can decrease the output voltage
to be below the OVP value or increase the OVP value to be greater than the
output voltage).
When [CH1|CH2|CH3] is omitted, the system clears the label of the OVP
occurred on the current channel.
[:SOURce[
to clear the circuit and label of the OVP occurred on the specified channel and
turn on the corresponding channel output.
query whether OVP occurred on the specified channel.
Example
:OUTP:OVP:QUES? CH1 /*Query whether overvoltage protection currently
occurred on CH1 and the query returns YES*/
(Operation: decrease the output voltage to be below the OVP value or increase the
OVP value to be greater than the output voltage)
:OUTP:OVP:CLEAR CH1 /*Clear the label of the overvoltage protection occurred
on CH1*/
:OUTP:OVP:QUES? CH1 /*Query whether OVP occurred on CH1 and the query
returns NO*/
Related
Commands
:OUTPut:OVP:ALAR?
:OUTPut:OVP:QUES?
[:SOURce[