Timer:templet:fallrate, Timer:templet:interval, Timer:templet:fallrate -134 – RIGOL DP832A User Manual
Page 158: Timer:templet:interval -134
RIGOL
2-134
DP800 Programming Guide
:TIMEr:TEMPlet:FALLRate
Syntax
:TIMEr:TEMPlet:FALLRate
:TIMEr:TEMPlet:FALLRate?
Description
Set the fall index of ExpFall.
Query the fall index of ExpFall.
Parameter
Name
Type
Range
Default
Integer
0 to 10
0
Explanation
When the templet currently selected is ExpFall (
), the timer
parameters created cannot reach the minimum due to the characteristic of the
exponential function. The range of the timer parameters created is related to the
fall index currently set. The larger the fall index is, the larger the range of the timer
parameters will be.
Return Format
The query returns an integer from 0 to 10, for example, 5.
Example
:TIME:TEMP:FALLR 5 /*Set the fall index of ExpFall to 5*/
:TIME:TEMP:FALLR? /*Query the fall index of ExpFall and the query returns 5*/
Related
Command
:TIMEr:TEMPlet:INTErval
Syntax
:TIMEr:TEMPlet:INTErval
:TIMEr:TEMPlet:INTErval?
Description
Set the time interval.
Query the current time interval.
Parameter
Name
Type
Range
Default
Integer
1s to 99999s
1s
Explanation
The time interval refers to the time required for the instrument to output each
group of timer parameters created using the templet currently selected.
The Pulse templet does not support this parameter.
Return Format
The query returns an integer from 1 to 99999, for example, 15.
Example
:TIME:TEMP:INTE 15 /*Set the time interval to 15s*/
:TIME:TEMP:INTE? /*Query the current time interval and the query returns 15*/