*PSC
Syntax
*PSC {0|1}
*PSC?
Description
Enable or disable the function to clear the enable registers of the status byte and
standard event registers at power-on.
Query the state of the function to clear the enable registers of the status byte and
standard event registers at power-on.
Parameter
Name
Type
Range
Default
{0|1}
Discrete
0|1
0
Explanation
*PSC 1 denotes clearing the enable registers of the status byte and standard
event registers at power-on; *PSC 0 denotes that the enable registers of the
status byte and standard event registers will not be affected at power-on.
clear the enable registers of the status byte and standard event registers
respectively.
Return Format
The query returns 0 or 1.
Example
*PSC 1 /*Enable the function to clear the enable registers of the status byte and
standard event registers at power-on*/
*PSC? /*Query the setting of the power-on status clear and the query returns 1*/
Related
Commands
*RCL
Syntax
*RCL {1|2|3|4|5|6|7|8|9|10}
Description
Recall the instrument state stored in the internal memory.
Parameter
Name
Type
Range
Default
{1|2|3|4|5|6|7|8|9|10}
Discrete
1|2|3|4|5|6|7|8|9|10
None
Explanation
The power supply provides 10 storage locations (numbered 1 to 10) for
instrument states. This command recalls the instrument state stored in the
specified location. Selecting number 1 to 10 can recall the instrument states
stored in the corresponding locations respectively.
This command is only available when a state file has been stored in the
specified storage location in the internal memory.
recall the instrument state stored in the internal memory.
Related