RIGOL DP832A User Manual
Page 34
RIGOL
2-10
DP800 Programming Guide
(multi-channel models) or current range (single-channel model) when only
[CH1|CH2|CH3] is specified, for example, CH1:8V/5A,5.000,1.0000.
The query returns the voltage setting value and current setting value of the
current channel (multi-channel models) or current range (single-channel
model) when all the parameters are omitted, for example, 5.000,1.0000.
Example
:APPL CH1,5,1 /*Set the voltage and current of CH1 to 5V and 1A respectively*/
:APPL? CH1 /*Query the voltage and current setting values of CH1 and the
query returns CH1:8V/5A,5.000,1.0000*/
Related
Commands
[:SOURce[
[:SOURce[
Note
[1]
: In this command system, commands that can use channel names (such as CH1 and CH2) as parameters can
also use channel ranges (such as P8V and P30V) as parameters unless otherwise noted. The corresponding relations of
the channel name parameters and channel range parameters supported by DP800 series are as follows.
DP831A/DP831: CH1---P8V; CH2---P30V; CH3---N30V
DP832A/DP832: CH1---P30V; CH2---P30V2; CH3---P5V
DP821A/DP821: CH1---P60V; CH2---P8V