Output commands, Output commands -52 – RIGOL DP832A User Manual
Page 76
RIGOL
2-52
DP800 Programming Guide
:OUTPut Commands
The :OUTPut commands are used to enable or disable the channel output, overvoltage/overcurrent
protection functions, track function and Sense function, query the output mode of the channel as well as set
and query the related information of overvoltage/overcurrent protection. For single-channel model, you can
also select and query the current range of the channel. The range and default value of
overvoltage/overcurrent protection corresponding to each channel (multi-channel models) or range
(single-channel model) of different models are as shown in the table below.
Table 2-2 Range and default value of overvoltage/overcurrent protection of each channel (range) of
different models of DP800 series
Channel (Range)
OVP/OCP Settable Range
OVP/OCP Default Value
DP831A
CH1 (8V/5A)
CH2 (30V/2A)
CH3 (-30V/2A)
1mV to 8.8V/0.1mA to 5.5A
1mV to 33V/0.1mA to 2.2A
-1mV to -33V/0.1mA to 2.2A
8.800V/5.5000A
33.000V/2.2000A
-33.000V/2.2000A
DP832A
CH1 (30V/3A)
CH2 (30V/3A)
CH3 (5V/3A)
1mV to 33V/1mA to 3.3A
1mV to 33V/1mA to 3.3A
1mV to 5.5V/1mA to 3.3A
33.000V/3.300A
33.000V/3.300A
5.500V/3.300A
DP821A
CH1 (60V/1A)
CH2 (8V/10A)
1mV to 66V/0.1mA to 1.1A
1mV to 8.8V/1mA to 11A
66.000V/1.1000A
8.800V/11.000A
DP811A
Range1 (20V/10A)
Range2 (40V/5A)
1mV to 22V/0.1mA to 11A
1mV to 44V/0.1mA to 5.5A
22.000V/11.0000A
DP831
[1]
CH1 (8V/5A)
CH2 (30V/2A)
CH3 (-30V/2A)
10mV to 8.8V/1mA to 5.5A
10mV to 33V/1mA to 2.2A
-10mV to -33V/1mA to 2.2A
8.800V/5.500A
33.00V/2.200A
-33.00V/2.200A
DP832
[1]
CH1 (30V/3A)
CH2 (30V/3A)
CH3 (5V/3A)
10mV to 33V/1mA to 3.3A
10mV to 33V/1mA to 3.3A
10mV to 5.5V/1mA to 3.3A
33.00V/3.300A
33.00V/3.300A
5.50V/3.300A
DP821
[1]
CH1 (60V/1A)
CH2 (8V/10A)
10mV to 66V/10mA to 1.1A
10mV to 8.8V/10mA to 11A
66.00V/1.100A
8.80V/11.00A
DP811
[1]
Range1 (20V/10A)
Range2 (40V/5A)
10mV to 22V/10mA to 11A
10mV to 44V/10mA to 5.5A
22.00V/11.00A
Note
[1]
: When DP831 (DP832, DP821 or DP811) is installed with the high resolution option, its overvoltage/overcurrent
protection settable range and default value of each channel are the same with those of DP831A (DP832A, DP821A or
DP811A).
Command List: