beautypg.com

Preset:key, Preset:key -69 – RIGOL DP832A User Manual

Page 93

background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-69

:PRESet:KEY

Syntax

:PRESet:KEY {DEFAULT|USER1|USER2|USER3|USER4}
:PRESet:KEY?

Description

Define the setting recalled by Preset at the front panel.

Query the setting recalled by Preset at the front panel.

Parameter

Name

Type

Range

Default

{DEFAULT|USER1|USER2|

USER3|USER4}

Discrete

DEFAULT|USER1|USER2|

USER3|USER4

DEFAULT

Explanation

 DP800 supports to restore the instrument to the default setting (DEFAULT) or

recall the user-defined setting (USER1, USER2, USER3 and USER4).

 After sending this command to define the setting recalled by Preset at the

front panel, sending the

:PRESet[:APPLy]

command or pressing Preset at the

front panel can restore the instrument to the default setting or recall the

specified user-defined setting.

 When user-defined setting is selected, you can use the :PRESet:USER:SET

series commands to set the parameters of the user-defined setting currently

selected.

Return Format

The query returns DEFAULT, USER1, USER2, USER3 or USER4.

Example

:PRES:KEY USER1 /*Define the setting recalled by Preset at the front panel to

the first group of user-defined setting*/

:PRES:KEY? /*Query the setting recalled by Preset at the front panel and

the query returns USER1*/

Related

Command

:PRESet[:APPLy]