Timer:templet:maxvalue, Timer:templet:maxvalue -136 – RIGOL DP832A User Manual
Page 160
RIGOL
2-136
DP800 Programming Guide
:TIMEr:TEMPlet:MAXValue
Syntax
:TIMEr:TEMPlet:MAXValue {
:TIMEr:TEMPlet:MAXValue? [MINimum|MAXimum]
Description
Set the maximum voltage or current of the templet currently selected.
Query the maximum voltage or current of the templet currently selected.
Parameter
Name
Type
Range
Default
Real
Voltage or current range of the
channel currently selected
Refer to the
"Explanation"
Explanation
This command sets the maximum voltage or current (depend on
command). When the editing object is voltage, this
command sets the maximum voltage; when the editing object is current, this
command sets the maximum current.
For the DP800 series (except CH3 (-30V/2A) of DP831A/DP831), the maximum
voltage and current of all the templets are 1V and 1A. For CH3 (-30V/2A) of
DP831A/DP831, the maximum voltage and current of all the templets are -1V
and 1A.
When the Pulse templet is used, this command is used to set the high level.
Return Format
The query returns the maximum voltage or current of the templet currently
selected, for example, 5.000 or 5.3000.
Example
:TIME:TEMP:OBJ V,2 /*Set the current editing object to voltage and set the
current to 2A*/
:TIME:TEMP:MAXV 5 /*Set the maximum voltage of the templet currently
selected to 5V*/
:TIME:TEMP:MAXV? /*Query the maximum voltage of the templet currently
selected and the query returns 5.000*/
Related