Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-51

:MONItor:VOLTage[:VALue]

Syntax

:MONItor:VOLTage[:VALue] {|MINimum|MAXimum}
:MONItor:VOLTage[:VALue]? [MINimum|MAXimum]

Description

Set the voltage value of the monitor condition (the current channel).
Query the voltage value of the monitor condition (the current channel).

Parameter

Name

Type

Range

Default

Real

Refer to the "Explanation"

Explanation

For multi-channel models, the range of is from 0 to the maximum voltage

of the current channel and the default is 0.5*rated voltage of the current channel;

for single-channel model, the range of is from 0 to the maximum voltage

of the current range and the default is 0.5*rated voltage of range 1.

Return Format

The query returns the voltage of the current monitor condition (the current

channel), for example, 5.000.

Example

:MONI:VOLT 5 /*Set the voltage of the monitor condition (the current channel)

to 5V*/

:MONI:VOLT? /*Query the voltage of the current monitor condition (the

current channel) and the query returns 5.000*/

Related

Command

:MONItor:VOLTage:CONDition