Source[]]:voltage:protection:clear, Source[]]:voltage:protection:clear -93 – RIGOL DP832A User Manual
Page 117
RIGOL
DP800 Programming Guide
2-93
[:SOURce[
Syntax
[:SOURce[
Description
Clear the circuit and label of the OVP occurred on the corresponding channel and
turn on the output of the corresponding channel.
Parameter
Name
Type
Range
Default
[
Integer
1|2|3
1
Explanation
[:SOURce[
to query whether OVP occurred on the specified channel.
Before executing the command, make sure that the reason that causes the
OVP on the specified channel is cleared (you can decrease the output voltage
to be below the OVP value or increase the OVP value to be greater than the
output voltage); otherwise, the command is invalid. Executing this command
will clear the circuit and label of the OVP occurred on the specified channel and
turn on the output of the corresponding channel.
When [:SOURce[
label of the OVP occurred on the current channel.
command to only clear the label of the
OVP occurred on the specified channel.
Example
:VOLT:PROT:TRIP? /*Query whether OCP occurred on the current channel
and the query returns YES*/
(Operation: decrease the output voltage to be below the OVP value or increase the
OVP value to be greater than the output voltage)
:VOLT:PROT:CLE /*Clear the circuit and label of the OVP occurred on the
current channel*/
:VOLT:PROT:TRIP? /*Query whether OVP occurred on the current channel
and the query returns NO*/
Related
Commands
[:SOURce[