DP800 Programming Guide
2-95
[:SOURce[
Syntax
[:SOURce[
[:SOURce[
Description
Enable or disable the overvoltage protection (OVP) function of the specified
channel.
Query the status of the overvoltage protection (OVP) function of the specified
channel.
Parameter
Name
Type
Range
Default
[
Integer
1|2|3
1
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
When the overvoltage protection function of the specified channel is enabled
[:SOURce[
), the output turns off
automatically when the output voltage exceeds the overvoltage protection limit
currently set. You can send the
[:SOURce[
overvoltage protection occurred on the specified channel currently.
When [:SOURce[
corresponding parameter of the channel currently selected.
command to enable or disable the
OVP function of the specified channel.
[:SOURce[
query the current OVP value of the specified channel.
Return Format
The query returns ON or OFF.
Example
:VOLT:PROT:STAT ON /*Enable the OVP function of the current channel*/
:VOLT:PROT:STAT? /*Query the status of the OVP function of the current
channel and the query returns ON*/
Related
Commands
[:SOURce[
:OUTPut:OVP[:STATe]
[:SOURce[