Output:ovp:alar, Output:ovp:ques, Output:ovp:alar? -58 – RIGOL DP832A User Manual
Page 82: Output:ovp:ques? -58, Output:ovp:alar? :output:ovp:ques
RIGOL
2-58
DP800 Programming Guide
:OUTPut:OVP:ALAR?
:OUTPut:OVP:QUES?
Syntax
:OUTPut:OVP:ALAR? [CH1|CH2|CH3]
:OUTPut:OVP:QUES? [CH1|CH2|CH3]
Description
Query whether OVP occurred on the specified channel.
Parameter
Name
Type
Range
Default
[CH1|CH2|CH3]
Discrete
CH1|CH2|CH3
None
Explanation
Overvoltage protection (OVP) refers to that the output turns off automatically
when the actual output voltage of the channel exceeds the OVP value.
When [CH1|CH2|CH3] is omitted, the system queries whether OVP occurred on
the current channel.
command to clear the OVP occurred on
the specified channel.
Return Format
The query returns YES or NO.
Example
:OUTP:OVP:ALAR? CH1
:OUTP:OVP:QUES? CH1 /*Query whether OVP occurred on CH1 and the query
returns YES*/
Related
Command