:SYSTem:SELF:TEST:FAN?

Syntax

:SYSTem:SELF:TEST:FAN?

Description

Query the self-test result of the fan.

Explanation

You can also send the

*TST?

command to query the self-test results (including the

self-test results of the TopBoard, BottomBoard and fan).

Return Format

The query returns PASS or FAIL.

Related

Command

*TST?

:SYSTem:SELF:TEST:TEMP?

Syntax

:SYSTem:SELF:TEST:TEMP?

Description

Query the self-test result of the temperature.

Return Format

The query returns the temperature value and the unit is ℃, for example, 23.67.

:SYSTem:TRACKMode

Syntax

:SYSTem:TRACKMode {SYNC|INDE}
:SYSTem:TRACKMode?

Description

Set the track mode.
Query the current track mode.

Parameter

Name

Type

Range

Default

{SYNC|INDE}

Discrete

SYNC|INDE

SYNC

Explanation

 The specified channels of DP831A, DP831, DP832A and DP832 support the track

function. You can select the desired track mode according to your need.

 SYNC: synchronous mode. For two channels (the channels should be of the

same instrument) that support the track function, the track function of the other

channel will be enabled or disabled at the same time when the track function of

a channel is enabled or disabled.

 INDE: independent mode. For two channels (the channels should be of the

same instrument) that support the track function, the status of the track

function of the other channel will not be affected when the track function of a

channel is enabled or disabled.

Return Format

The query returns SYNC or INDE.

Example

:SYST:TRACKM SYNC /*Set the track mode to synchronous*/
:SYST:TRACKM? /*Query the current track mode and the query returns

SYNC*/

Related

Command

:SYSTem:ONOFFSync