Timer:templet:invert, Timer:templet:invert -135 – RIGOL DP832A User Manual
Page 159
RIGOL
DP800 Programming Guide
2-135
:TIMEr:TEMPlet:INVErt
Syntax
:TIMEr:TEMPlet:INVErt {ON|OFF}
:TIMEr:TEMPlet:INVErt?
Description
Enable or disable the invert function of the templet currently selected.
Query whether the invert function of the templet currently selected is enabled.
Parameter
Name
Type
Range
Default
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
When the invert function is enabled, the instrument will first turn the preset
waveform upside down and then create timer parameters.
Only the Sine, Pulse and Ramp templets support this function.
Return Format
The query returns ON or OFF.
Example
:TIME:TEMP:INVE ON /*Enable the invert of the templet currently selected*/
:TIME:TEMP:INVE? /*Query whether the invert of the templet currently
selected is enabled and the query returns ON*/