:PRESet:USER[]:SET:CURRent

Syntax

:PRESet:USER[]:SET:CURRent {|MINimum|MAXimum}
:PRESet:USER[]:SET:CURRent? [MINimum|MAXimum]

Description

Set the current value of the current channel in the specified user-defined setting.
Query the current value of the current channel in the specified user-defined setting.

Parameter

Name

Type

Range

Default

[]

Discrete

1|2|3|4

None

Real

Current range of the current channel Refer to Table 2-3

Explanation

 When [] is 1, 2, 3 or 4, the system sets the related parameters of the first,

second, third or fourth group of user-defined setting respectively. When []

is omitted, the system sets the parameters of the user-defined setting

currently selected by default.

 For the current range of the current channel, please refer to Table 2-1.

 By default, this command sets the parameter of the channel currently selected.

For multi-channel models, if you want to set the corresponding parameters of

other channels, please use the

:INSTrument:NSELect

,

:INSTrument[:SELEct]

or

:INSTrument[:SELect]

command to set the desired channel as the current

channel.

 You can send the

:PRESet:USER[]:SET:DEFault

command to restore the

parameters of the specified user-defined setting to the default values.

Return Format

The query returns the current value, for example, 1.5000.

Example

:PRES:USER1:SET:CURR 1.5 /*Set the current of the current channel in the first

group of user-defined setting to 1.5A*/

:PRES:USER1:SET:CURR? /*Query the current of the current channel in the first

group of user-defined setting and the query returns

1.5000*/

Related

Commands

:INSTrument:NSELect

:INSTrument[:SELEct]
:INSTrument[:SELect]

:PRESet:USER[]:SET:DEFault