Command List:

:STATus:QUEStionable:CONDition?

:STATus:QUEStionable:ENABle

:STATus:QUEStionable[:EVENt]?

:STATus:QUEStionable:INSTrument:ENABle

:STATus:QUEStionable:INSTrument[:EVENt]?

:STATus:QUEStionable:INSTrument:ISUMmary[]:COND?

:STATus:QUEStionable:INSTrument:ISUMmary[]:ENABle

:STATus:QUEStionable:INSTrument:ISUMmary[][:EVENt]?

:STATus:QUEStionable:CONDition?

Syntax

:STATus:QUEStionable:CONDition?

Description

Query the condition register of the questionable status register.

Explanation

 This command is only applicable to the single-channel models.

 This command returns a decimal value corresponding to the sum of the binary

weights of all the bits in the register. For the definitions of the bits in the

questionable status register and their corresponding decimal values, refer to

Table 1-4.

For example, when the query returns 0, the power supply is turned off or in the

unregulated mode; when the query returns 1, the power supply is in CC mode;

when the query returns 2, the power supply is in CV mode; when the query

returns 3, power supply failure occurs.

Return Format

The query returns a decimal value corresponding to the sum of the binary weights

of all the bits in the register, for example, 1.

Example

:STAT:QUES:COND? /*Query the condition register of the questionable status

register and the query returns 1*/