Status commands, Status:questionable:condition, Status commands -98 – RIGOL DP832A User Manual
Page 122: Status:questionable:condition? -98
RIGOL
2-98
DP800 Programming Guide
:STATus Commands
Command List:
:STATus:QUEStionable:CONDition?
:STATus:QUEStionable[:EVENt]?
:STATus:QUEStionable:INSTrument:ENABle
:STATus:QUEStionable:INSTrument[:EVENt]?
:STATus:QUEStionable:INSTrument:ISUMmary[
:STATus:QUEStionable:INSTrument:ISUMmary[
:STATus:QUEStionable:INSTrument:ISUMmary[
:STATus:QUEStionable:CONDition?
Syntax
:STATus:QUEStionable:CONDition?
Description
Query the condition register of the questionable status register.
Explanation
This command is only applicable to the single-channel models.
This command returns a decimal value corresponding to the sum of the binary
weights of all the bits in the register. For the definitions of the bits in the
questionable status register and their corresponding decimal values, refer to
For example, when the query returns 0, the power supply is turned off or in the
unregulated mode; when the query returns 1, the power supply is in CC mode;
when the query returns 2, the power supply is in CV mode; when the query
returns 3, power supply failure occurs.
Return Format
The query returns a decimal value corresponding to the sum of the binary weights
of all the bits in the register, for example, 1.
Example
:STAT:QUES:COND? /*Query the condition register of the questionable status
register and the query returns 1*/