Monitor:stopway, Monitor:stopway -49 – RIGOL DP832A User Manual
Page 73
RIGOL
DP800 Programming Guide
2-49
:MONItor:STOPway
Syntax
:MONItor:STOPway {OUTOFF|WARN|BEEPER|NONE},{ON|OFF}
:MONItor:STOPway?
Description
Set the stop mode of the monitor (the current channel).
Query the stop mode of the monitor (the current channel).
Parameter
Name
Type
Range
Default
{OUTOFF|WARN|BEEPER}
Discrete OUTOFF|WARN|BEEPER
None
{ON|OFF}
Bool
ON|OFF
ON
Explanation
The monitor can monitor the current output state of the instrument. When the
monitor is enabled (
) and the channel output state meets the
monitor condition set, the instrument will turn off the output, display the
corresponding prompt message or sound the beeper according to the stop mode
selected (you can select one or more stop modes).
Return Format
The query returns the on/off status of the three stop modes separated by commas,
for example, OutputOff:ON,Warn:ON,Beep:ON.
Example
:MONI:STOP OUTOFF,ON /*Enable the "OutpOff" stop mode*/
:MONI:STOP WARN,ON /*Enable the "Warning" stop mode*/
:MONI:STOP BEEPER,OFF /*Disable the "Beeper" stop mode*/
:MONI:STOP? /*Query the current stop modes of the monitor and
the query returns
OutputOff:ON,Warn:ON,Beep:OFF*/
Related
Command