Output:ocp[:state, Output:ocp[:state] -56 – RIGOL DP832A User Manual
Page 80
RIGOL
2-56
DP800 Programming Guide
:OUTPut:OCP[:STATe]
Syntax
:OUTPut:OCP[:STATe] [CH1|CH2|CH3,]{ON|OFF}
:OUTPut:OCP[:STATe]? [CH1|CH2|CH3]
Description
Enable or disable the overcurrent protection (OCP) function of the specified
channel.
Query the status of the overcurrent protection (OCP) function of the specified
channel.
Parameter
Name
Type
Range
Default
[CH1|CH2|CH3]
Discrete
CH1|CH2|CH3
None
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
When the overcurrent protection function is enabled, the output will turn off
automatically when the output current exceeds the overcurrent protection
value currently set (
). You can send
overcurrent protection occurred on the specified channel currently.
When [CH1|CH2|CH3] is omitted, the system executes the corresponding
operation on the current channel.
Return Format
The query returns ON or OFF.
Example
:OUTP:OCP CH1,ON /*Enable the overcurrent protection function of CH1*/
:OUTP:OCP? CH1 /*Query the status of the overcurrent protection function of
CH1 and the query returns ON*/
Related
Commands
:OUTPut:OCP:ALAR?
:OUTPut:OCP:QUES?