Source[]]:voltage:range, Source[]]:voltage:range -97 – RIGOL DP832A User Manual
Page 121
RIGOL
DP800 Programming Guide
2-97
[:SOURce[
Syntax
[:SOURce[
[:SOURce[
Description
Select the range of the channel.
Query the range currently selected of the channel.
Parameter
Name
Type
Range
Default
[
Integer
1
1
{P20V|P40V|LOW|HIGH}
Discrete
P20V|P40V|LOW|HIGH
None
Explanation
This command is only applicable to single-channel model (DP811A/DP811).
[:SOURce[
P20V or LOW represents Range1: 20V/10A and P40V or HIGH represents
Range2: 40V/5A.
command to select the current range.
Return Format
The query returns the rated voltage and current of the range selected, for example,
20V/10A.
Example
:VOLT:RANG P20V /*Select Range1: 20V/10A as the current range*/
:VOLT:RANG? /*Query the range currently selected and the query returns
20V/10A*/
:VOLT:RANG HIGH /*Select Range2: 40V/5A as the current range*/
:VOLT:RANG? /*Query the range currently selected and the query returns
40V/5A*/