Output:sense, Output[:state, Output:sense -63 – RIGOL DP832A User Manual
Page 87: Output[:state] -63
RIGOL
DP800 Programming Guide
2-63
:OUTPut:SENSe
Syntax
:OUTPut:SENSe [CH1|CH2|CH3,]{ON|OFF}
:OUTPut:SENSe? [CH1|CH2|CH3]
Description
Enable or disable the Sense function of the channel.
Query the status of the Sense function of the channel.
Parameter
Name
Type
Range
Default
[CH1|CH2|CH3]
Discrete
CH1|CH2|CH3
None
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
In large current output, to ensure that the load can acquire accurate voltage
drop, CH2 of DP821A/DP821 and DP811A/DP811 provide Sense (remote
sense) working mode. In this mode, the voltage at the load terminal instead of
the output voltage of the power supply is measured to compensate the voltage
drop caused by the load lead automatically.
This command is only applicable to the channels (DP821A/DP821 (CH2) and
DP811A/DP811) that support the Sense function. For channels that do not
support the Sense function, the query returns NONE.
DP821A/DP821/DP811A/DP811 provides output terminal on the front panel for
the Sense working mode. Besides, the output interface on the rear panel of
DP811A/DP811 can also be used for the Sense working mode. But, only one
output mode can be selected at one time and the output terminal on the front
panel can provide higher precision than the output interface on the rear panel.
Return Format
The query returns ON or OFF. For channels that do not support the Sense function,
the query returns NONE.
Example
:OUTP:SENS CH1,ON /*Enable the Sense function of the DP811A channel*/
:OUTP:SENS? CH1 /*Query the status of the Sense function of the DP811A
channel and the query returns ON*/
:OUTPut[:STATe]
Syntax
:OUTPut[:STATe] [CH1|CH2|CH3,]{ON|OFF}
:OUTPut[:STATe]? [CH1|CH2|CH3]
Description
Enable or disable the output of the specified channel.
Query the output status of the specified channel.
Parameter
Name
Type
Range
Default
[CH1|CH2|CH3]
Discrete
CH1|CH2|CH3
None
{ON|OFF}
Bool
ON|OFF
OFF
Explanation
Make sure that the current setting will not affect the device connected to the
power supply before enabling the channel output.
When [CH1|CH2|CH3] is omitted, the system performs the corresponding
operation on the current channel.
Return Format
The query returns ON or OFF.
Example
:OUTP CH1,ON /*Enable the output of CH1*/
:OUTP? CH1 /*Query the current output status of CH1 and the query returns
ON*/