86

Alarms and alerts

Welch Allyn Propaq LT Vital Signs Monitor

To respond at Acuity to an alert

For a monitor in communication with Acuity, either the monitor or Acuity can detect
alarms and alerts. The alarms are displayed in both places. See Acuity Directions For use.

About battery charge status

Table 6. Battery status indicators

Display

Status/alert
text

Battery

Monitor

Recommended
action

Green

Fully charged

Okay

None

Green

Partially full

Okay

None

Green

Partially full; charging Okay

None

Yellow

EQUIPMENT ALERT
Low Battery.
Charge battery soon.

Low

Can function for up to
30 minutes, but NIBP
is disabled.

Prepare to discontinue
monitoring. If possible,
insert the monitor into a
cradle.

Yellow

Low; charging

Normal function.

Do not remove the
monitor from the cradle.

Red

EQUIPMENT ALERT
Battery Too Low.
Shutting down.

Almost completely
discharged

Shutting down soon.
Can function for up to
5 minutes, but NIBP is
disabled.

Prepare to discontinue
monitoring. If possible,
insert the monitor into a
cradle.
If this indicator appears
when the monitor is in a
powered cradle, then
the battery is damaged
and must be replaced.
In this case, all stored
patient data will be
deleted when the
monitor is removed
from the cradle.

Red

Almost completely
discharged; charging

Normal function.

Do not remove the
monitor from the cradle.