About battery charge status – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 90

86
Alarms and alerts
Welch Allyn Propaq LT Vital Signs Monitor
To respond at Acuity to an alert
For a monitor in communication with Acuity, either the monitor or Acuity can detect
alarms and alerts. The alarms are displayed in both places. See Acuity Directions For use.
About battery charge status
Table 6. Battery status indicators
Display
Status/alert
text
Battery
Monitor
Recommended
action
Green
Fully charged
Okay
None
Green
Partially full
Okay
None
Green
Partially full; charging Okay
None
Yellow
EQUIPMENT ALERT
Low Battery.
Charge battery soon.
Low
Can function for up to
30 minutes, but NIBP
is disabled.
Prepare to discontinue
monitoring. If possible,
insert the monitor into a
cradle.
Yellow
Low; charging
Normal function.
Do not remove the
monitor from the cradle.
Red
EQUIPMENT ALERT
Battery Too Low.
Shutting down.
Almost completely
discharged
Shutting down soon.
Can function for up to
5 minutes, but NIBP is
disabled.
Prepare to discontinue
monitoring. If possible,
insert the monitor into a
cradle.
If this indicator appears
when the monitor is in a
powered cradle, then
the battery is damaged
and must be replaced.
In this case, all stored
patient data will be
deleted when the
monitor is removed
from the cradle.
Red
Almost completely
discharged; charging
Normal function.
Do not remove the
monitor from the cradle.