beautypg.com

Default conditions, Audible alarms suspension time default, Nibp units default – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 131: Pacer indicator on default, Display all ecg vectors

background image

Directions for Use

127

Audible Alarms Suspension Time Default

__ Disable (alarm tones cannot be suspended)
__ 90 seconds
__ 2 minutes
__ 3 minutes
__ 4 minutes
__ 5 minutes
__ 10 minutes
__ 15 minutes
__ 30 minutes
__ 60 minutes
__ Always On (alarm tones always suspended)

While monitoring a patient, the clinician can use the settings menu to suspend all alarm
tones for all parameters for a configurable period.

NIBP Units Default

__ mmHg
__ kPa

Default Conditions

Pacer Indicator On Default

(Yes)
(No)

If a patient being monitored has a pacemaker, the monitor detects the occurrence of
pacemaker signals and optionally indicates them with vertical dashed lines in the
waveform.

If the pacemaker signal is strong enough, the monitor displays it as a spike. This spike is
displayed whether the pacemaker indicator is on or off.

Display All ECG Vectors

(Yes)
(No)

Note

This parameter does not affect the behavior of

, (Silence/Reset). Regardless

of how this parameter is configured, pressing

silences a sounding alarm

tone for 90 seconds.

Suspend can be turned on or off by the clinician, but the suspend period
cannot be changed by the clinician.

Note

If the monitor is connected to an Acuity Central Station, this is forced to mmHg.

HR

18:45:28

III

1

mV

cm

MON

80

Pacemaker signal indicators