Changing the default settings, Taking nibp readings using turbo mode, Nibp measurements in power-saving mode – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 73: Nibp disabled when the battery is low

Directions for Use
Standalone monitoring
69
To stop automatic NIBP readings
1.
Highlight NIBP and press
.
2.
Highlight NIBP Mode and select Manual.
Taking NIBP readings using Turbo mode
In Turbo mode, the monitor starts an NIBP reading and then takes as many more readings
as possible within five minutes.
To use the Turbo mode
1.
Select and apply the NIBP cuff and hose. (
“Monitoring blood pressure (NIBP)”
2.
Highlight NIBP and press
.
3.
Highlight NIBP Mode and select Turbo.
To end the Turbo mode
Press
or select NIBP Mode Manual. The monitor returns to Auto NIBP
measurement mode.
NIBP measurements in power-saving mode
When a manual or turbo NIBP activity awakens the monitor from power-saving mode
(
on page 44), cuff inflation pressure is reset to default levels and cuff
inflation is delayed for up to 8 seconds.
NIBP disabled when the battery is low
If the battery is low and the monitor is operating on battery power, NIBP functions are
disabled and the monitor displays the message “NIBP off. Low battery.”
If you attempt to start an NIBP measurement during a low-battery condition, the monitor
displays an equipment alert with the message “Low battery. NIBP disabled.”
Changing the default settings
You can change the default settings (the monitor configuration) using the following
methods:
•
Save the current settings as the new default settings.
•
Restore the last downloaded configuration.
•
Restore the factory settings.
Note
If you cycle the monitor power, NIBP returns to auto mode.
Note
Inserting a monitor into a powered cradle during a low-battery condition
immediately enables NIBP monitoring.