Directions for Use

Standalone monitoring

69

To stop automatic NIBP readings

1.

Highlight NIBP and press

.

2.

Highlight NIBP Mode and select Manual.

Taking NIBP readings using Turbo mode

In Turbo mode, the monitor starts an NIBP reading and then takes as many more readings
as possible within five minutes.

To use the Turbo mode

1.

Select and apply the NIBP cuff and hose. (

“Monitoring blood pressure (NIBP)”

on

page 65.)

2.

Highlight NIBP and press

.

3.

Highlight NIBP Mode and select Turbo.

To end the Turbo mode

Press

or select NIBP Mode Manual. The monitor returns to Auto NIBP

measurement mode.

NIBP measurements in power-saving mode

When a manual or turbo NIBP activity awakens the monitor from power-saving mode
(

“Power saving”

on page 44), cuff inflation pressure is reset to default levels and cuff

inflation is delayed for up to 8 seconds.

NIBP disabled when the battery is low

If the battery is low and the monitor is operating on battery power, NIBP functions are
disabled and the monitor displays the message “NIBP off. Low battery.”

If you attempt to start an NIBP measurement during a low-battery condition, the monitor
displays an equipment alert with the message “Low battery. NIBP disabled.”

Changing the default settings

You can change the default settings (the monitor configuration) using the following
methods:

Save the current settings as the new default settings.

Restore the last downloaded configuration.

Restore the factory settings.

Note

If you cycle the monitor power, NIBP returns to auto mode.

Note

Inserting a monitor into a powered cradle during a low-battery condition