Specifications

Welch Allyn Propaq LT Vital Signs Monitor

NIBP

Characteristic

Specification

Method

Oscillometric

Regulatory standards

Meets AAMI SP10:2002, IEC 60601

-

1:1990, IEC 60601

-

2

-

30:1999(E);

EN 1060

-

1:1996, EN 1060

-

3:1997

Control

Automatic (at selected intervals), turbo, and manual

Auto intervals

1, 2, 3, 5, 10, 15, 30, and 60 minutes

Smartcuf

Available when ECG is used

Reported pressures

Systolic, diastolic, and mean plus real

-

time manometer pressure

Pressure resolution

1 mmHg

Systolic range

Alarm limits

Adult

30 to 260 mmHg

Pediatric

30 to 160 mmHg

Neonate

25 to 120 mmHg

Adult

low 30 to 258 mmHg

high

32 to 260 mmHg

Pediatric

low 30 to 158 mmHg

high

32 to 160 mmHg

Neonate

low 25 to 118 mmHg

high

27 to 120 mmHg

Diastolic range

Alarm limits

Adult

20 to 235 mmHg

Pediatric

15 to 130 mmHg

Neonate

10 to 105 mmHg}

Adult

low 20 to 233 mmHg

high

22 to 235 mmHg

Pediatric

low 15 to 128 mmHg

high

17 to 130 mmHg

Neonate

low 10 to 103 mmHg

high

12 to 105 mmHg

Mean range

Alarm limits

Adult

20 to 255 mmHg

Pediatric

15 to 140 mmHg

Neonate

10 to 110 mmHg

Adult

low 20 to 253 mmHg

high

22 to 255 mmHg

Pediatric

low 15 to 138 mmHg

high

17 to 140 mmHg

Neonate

low 10 to 108 mmHg

high

12 to 110 mmHg

Pulse rate range

Alarm limits

30 to 220 beats/min, in adult, pediatric, and neonate modes

Low

25 to 298 beats/min

high

27 to 300 beats/min

Static manometer accuracy

0

-

300mmHg ±3 mmHg or 2% of reading, whichever is greater. Can be checked

in the field, but no adjustments are provided (or needed).

Manometer baselining (zeroing)

Automatic

Atmospheric pressure compensation

Automatic

Minimum inflation pressure

Adult

100 mmHg

Pediatric

80 mmHg

Neonate

50 mmHg

Maximum inflation pressure

Adult

270 mmHg

Pediatric

170 mmHg

Neonate

132 mmHg

Default inflation pressure

Adult

160 mmHg

Pediatric

120 mmHg

Neonate

90 mmHg