Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 113

Directions for Use
Monitor configuration
109
HR/PR adult
Upper Alarm Limit
120 beats/min
27 - 300 beats/min
Monitor
Lower Alarm Limit
50 beats/min
25 - 298 beats/min
Monitor
HR/PR pediatric
Upper Alarm Limit
150 beats/min
27 - 300 beats/min
Monitor
Lower Alarm Limit
50 beats/min
25 - 298 beats/min
Monitor
HR/PR neonate
Upper Alarm Limit
200 beats/min
27 - 300 beats/min
Monitor
Lower Alarm Limit
100 beats/min
25 - 298 beats/min
Monitor
NIBP general
ParamSet for Upper NIBP
Systolic
15%
5% - 25%
Configuration Utility
ParamSet for Lower NIBP
Systolic
15%
5% - 25%
Configuration Utility
ParamSet for Upper NIBP
Diastolic
15%
5% - 25%
Configuration Utility
ParamSet for Lower NIBP
Diastolic
15%
5% - 25%
Configuration Utility
ParamSet for Upper NIBP
Mean
10%
5% - 15%
Configuration Utility
ParamSet for Lower NIBP
Mean
10%
5% - 15%
Configuration Utility
NIBP adult
Auto Interval (minutes)
15
1, 2, 3, 5, 10, 15, 30, 60
Monitor, Configuration
Utility
Mode
Manual
Manual, Auto
Monitor, Acuity
Smartcuf
Off
On, Off
Monitor, Acuity,
Configuration Utility
Systolic Upper Alarm Limit
220 mmHg
29.3 kPa
32 - 260 mmHg
4.3 - 34.7 kPa
Monitor, Acuity
Systolic Lower Alarm Limit
75 mmHg
10.0 kPa
30 - 258 mmHg
4.0 - 34.4 kPa
Monitor, Acuity
Diastolic Upper Alarm Limit
110 mmHg
14.7 kPa
22 - 235 mmHg
2.9 - 31.3 kPa
Monitor, Acuity
Diastolic Lower Alarm Limit
35 mmHg
4.7 kPa
20 - 233 mmHg
2.7 - 31.1 kPa
Monitor, Acuity
Mean Upper Alarm Limit
120 mmHg
16.0 kPa
22 - 255 mmHg
2.9 - 34.0 kPa
Monitor, Acuity
Mean Lower Alarm Limit
50 mmHg
6.7 kPa
20 - 253 mmHg
2.7 - 33.7 kPa
Monitor, Acuity
Table 10. Parameter configuration matrix (continued)
Parameter
Factory setting
Possible values
Where set